The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that per complaint details, ¿the augment wouldn¿t tighten down completely¿ during 2 separate procedures this week which resulted in a modified surgical procedure by ¿cementing the augment on both times¿.Reportedly, the surgeon performs many revisions and has not experienced this issue in the past.Responses to requested information/documentation was not provided; therefore, the root cause of the reported event could not be concluded.The patient impact beyond the modified surgical procedure could not be determined as the patient outcomes were not provided, although no patient injury or surgical delay was alleged.No further medical assessment is warranted at this time.Should additional clinically relevant documentation/information become available, the clinical/medical task may be re-evaluated.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Factors and/or some potential probable causes that could contribute to the reported event have been identified as but not limited to abnormal motion over time, bone degeneration, design of device, fit/sizing, lack of ingrowth, lifetime of device, and traumatic injury.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|