MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC SOLERA VOYAGER 4.75; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC SOLERA VOYAGER 4.75; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 6642002
Medical Device Problem Codes Material Deformation (2976); Mechanics Altered (2984)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via a medtronic representative regarding a patient implanted with a set-screw using drivers for an olif, flipped posterior for perc screws.It was reported that the set screw drivers were all stripped and were not allowing for set screw to break off.They had several pans sterile that allowed them to finish the case via the back up pans.All 4 instruments were stripped and unable to break off the set screws.The patient is doing fine, they had several back up pans allowing them to finish the case.The patient was in contact with the instruments intra procedure, and there was a 25 min.Addition to the procedure because they had to go down to spd for their back up pans several times.There were no patient symptoms or complications as a result of the event.All drivers were replaced by medtronic products.No further complications were reported/ anticipated.
 
Additional Manufacturer Narrative
H3: product analysis part# 6642002, lot# em16k037- visual and optical examination analysis revealed that the locking tabs have been bent inwards of the tip of the driver which will not allow the center shaft of the instrument to function and the retaining pin has been bent and the hex edges of the compressor have been rounded.This is consistent with bend stress overload.H6: updated patient code, eval.Code method, eval.Code result post analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLERA VOYAGER 4.75
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key10668659
Report Number1030489-2020-01419
Device Sequence Number16798794
Product Code LXH
UDI-Device Identifier00643169492455
UDI-Public00643169492455
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 01/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6642002
Device Catalogue Number6642002
Device Lot NumberEM16K037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer01/08/2021
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date01/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-