| Model Number |
7205357 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issues were observed.A functional evaluation revealed a jammed motor/blade stall error.The complaint was confirmed and the root cause has been associated with a mechanical component failure.A review of manufacturing records found the device has been in service for over 5 years, thus any malfunction presented at this point in time would not be related to the manufacture of the product.A complaint history review concluded this was a repeat issue.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that the motor drive unit was not turning on.The incident occurred before the procedure; no complications were reported.Results of investigation have concluded that this unit had a jammed motor/ blade stall which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|