• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 7 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CR FB INSRT SZ 7 8MM; ATTUNE IMPLANT : KNEE TIBIAL INSERT Back to Search Results
Model Number 1516-20-708
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Muscular Rigidity (1968); No Code Available (3191)
Date of Event 09/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address loosening of the tibial component at the cement to implant interface.Cement manufacturer was unknown.It was also reported that the patient was complaining of stiffness.Patient was scheduled as a poly swap vs femoral revision vs full revision.Poly was extracted.Femur was tapped on to check fixation and it appeared well fixed.Tibia was also tapped on to check fixation and came out with a few taps with a mallet on the underside of the component.Femur was then also extracted and an attune revision fb cemented stem component was implanted.Patella was retained.I was only able to get the lot number off of the tibial tray.Doi: unknown, dor: (b)(6) 2020, right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CR FB INSRT SZ 7 8MM
Common Device Name
ATTUNE IMPLANT : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10670901
Report Number1818910-2020-22178
Device Sequence Number19473375
Product Code JWH
UDI-Device Identifier10603295045915
UDI-Public10603295045915
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K101433
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1516-20-708
Device Catalogue Number151620708
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/28/2020
Supplement Date Received by Manufacturer11/13/2020
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date11/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CR FEM RT SZ 7 CEM; ATTUNE FB TIB BASE SZ 7 CEM; UNK ATTUNE KNEE PATELLA; ATTUNE CR FEM RT SZ 7 CEM; ATTUNE FB TIB BASE SZ 7 CEM; UNK ATTUNE KNEE PATELLA
Outcome Attributed to Adverse Event Required Intervention;
-
-