MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE PRIME; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE PRIME; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1011710-38
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Intimal Dissection (1333)
Event Date 09/03/2020
Event Type  Injury  
Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.The reported patient effect of intimal dissection is listed in the xience prime / xience prime small vessel (sv), and xience prime long length (ll), everolimus eluting coronary stent system (eecss), instructions for use as a known patient effect of coronary stenting procedures.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event Description
It was reported that the procedure was treat a mildly calcified de novo left circumflex (lcx) artery that is 95% stenosed.The lesion was pre-dilatated with a 1.5x12mm non-abbott balloon and a 3.5x38mm xience prime was successfully deployed at 10 atmospheres.Post dilatation with 3.5x8mm nc balloon at 14 atmospheres revealed a distal edge dissection.The dissection was treated with a 3.0x38mm xience prime.There were no adverse patient sequela and no clinically significant delay.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE PRIME
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10671002
MDR Text Key211094387
Report Number2024168-2020-08536
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/21/2022
Device Catalogue Number1011710-38
Device Lot Number9051041
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
3.5X8MM NC BALLOON
Patient Outcome(s) Required Intervention;
Patient Age49 YR
-
-