MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number 3542001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Contracture (1761); No Code Available (3191)
Date of Event 01/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Analysis was completed on a photo received of the suspect medical device.The physical product was not returned to mentor.Device evaluation summary: according to the information reported, the patient developed capsular contracture.During visual evaluation of the picture, no apparent damage or visual anomalies were observed in the product.Mentor concluded that the capsular contracture in the patient's breast was the result of the body's individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.Although the product was not received, the manufacturing records of the lot-serial number provided were identified.The manufacturing records were reviewed, and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Capsular contracture complaint information is consistently analyzed and monitored by mentor quality assurance to determine when further action is necessary.Reason for device explant and/or reoperation: bilateral breast capsular contracture.(b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old female patient who underwent a breast augmentation revision with mentor memorygel breast implant 200cc gel breast prostheses developed baker grade iii capsular contracture in both of her breasts.As a result, the patient underwent bilateral explantation and replacement with unspecified mentor gel breast prostheses on (b)(6) 2020.This medwatch report is for the 2nd of two gel breast prostheses.
 
Additional Manufacturer Narrative
On 02-dec-2020, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information reported, the patient developed capsular contracture on the breast.During the visual evaluation, a tear in the anterior view was observed measuring approximately 1.3 cm.Mentor concluded that the capsular contracture in the patient¿s breast was the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of our quality process, the manufacturing records of this lot-serial number were reviewed and the manufacturing standards were met prior to the release of this lot.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Complaint information is consistently analyzed and monitored by quality assurance to determine when further action is necessary.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 02-nov-2020, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key10671509
Report Number1645337-2020-12963
Device Sequence Number14134967
Product Code FTR
UDI-Device Identifier00081317000822
UDI-Public00081317000822
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2015
Device Model Number3542001
Device Catalogue Number3542001
Device Lot Number5997128
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Initial Date Received by Manufacturer 09/21/2020
Supplement Date Received by Manufacturer11/02/2020
12/02/2020
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date11/19/2020
12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
-
-