SMITH & NEPHEW, INC. PCH BSKT DCKBL UPBTR 1.5MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 012014
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/21/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the punch basket duckling was broken.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection revealed the front end assembly had separated from the handle assembly.The set screw and the actuator were missing.The device was unable to be functionally evaluated due to the fractured front end assembly.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated issue.The complaint was confirmed.Factors that could have contributed to the reported event include the application of excessive force resulting in instrument failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PCH BSKT DCKBL UPBTR 1.5MM
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10672014
Report Number1219602-2020-01577
Device Sequence Number2881672
Product Code NBH
UDI-Device Identifier03596010023360
UDI-Public03596010023360
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number012014
Device Catalogue Number012014
Device Lot Number50875210
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/21/2020
Supplement Date Received by Manufacturer09/21/2020
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date01/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-