| Model Number |
1011-0141-03 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the associated defibrillator failed to discharge using these internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The internal handles were returned to zoll medical australia and the customer's report was observed and attributed to the discharge button.The discharge button was caught in the "delivered (depression)" position, which prevented the handles from discharging.Analysis of reports of this type has not identified an increase in trend.
|
| |
|