• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION ASSY, MAIH, 2.7 DIA, 7 FT, W/SWITCH, 100-240V, PAC; INTERNAL HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION ASSY, MAIH, 2.7 DIA, 7 FT, W/SWITCH, 100-240V, PAC; INTERNAL HANDLE Back to Search Results
Model Number 1011-0141-03
Medical Device Problem Code Failure to Deliver Shock/Stimulation (1133)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that during biomed testing, the associated defibrillator failed to discharge using these internal handles.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The internal handles were returned to zoll medical australia and the customer's report was observed and attributed to the discharge button.The discharge button was caught in the "delivered (depression)" position, which prevented the handles from discharging.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSY, MAIH, 2.7 DIA, 7 FT, W/SWITCH, 100-240V, PAC
Common Device Name
INTERNAL HANDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key10672051
Report Number1220908-2020-03187
Device Sequence Number16830319
Product Code LDD
UDI-Device Identifier00847946022839
UDI-Public00847946022839
Combination Product (Y/N)N
Initial Reporter CountryAU
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1011-0141-03
Device Catalogue Number1011-0141-03
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/22/2020
Supplement Date Received by Manufacturer09/22/2020
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date11/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-