MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; LASER THERAPY PRODUCT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; LASER THERAPY PRODUCT Back to Search Results
Model Number PED-475-25
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/09/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the distal segment of the pipeline failed to open.The tip of the stent was repeatedly pushed and pulled, but still could not be opened.It was noted the stent was not positioned in a bend, more than 50% of the stent was deployed, re-sheathing was performed more than twice, and no additional steps or devices were used in an attempt to open the pipeline.A replacement stent was used to complete the procedure.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.Post-procedure angiography showed slowed blood flow.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm of the cavernous sinus with a max diameter of 5.87 mm and a 4.9 mm neck diameter.It was noted the patient's vessel tortuosity was moderate.Dual antiplatelet therapy (dapt) was ad ministered, and the platelet reactivity units (pru) level was up to standard.Medical history includes: hypertension, hyperglycemia (diabetes) and hyperlipidemia.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The pushwire was found to be kinked at 143.7cm from proximal end.The hypotube was found to be intact and the ptfe pulled back.The proximal bumper, re-sheathing marker and re-sheathing pad were found to be intact.The dps sleeves were found to be intact.The tip coil was found to be damaged.Due to the condition in which the braid was returned the distal and proximal ends could not be determined.Both ends of the pipeline flex braid were found fully collapsed and frayed.No other damages or anomalies were found with the device.Based on the device analysis and reported information, the customer¿s report of ¿failure/incomplete open distal¿ was confirmed as both ends of the pipeline flex braid were found to be collapsed and frayed.The customer reported having re-sheathed the device more than 2 times.The customer also reported the vessel tortuosity was moderate.It is likely the failure to open is the result of vessel tortuosity or re-sheathing the device more than the recommended two times.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
LASER THERAPY PRODUCT
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key10673058
Report Number2029214-2020-01025
Device Sequence Number16826350
Product Code OUT
UDI-Device Identifier00836462015883
UDI-Public00836462015883
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2022
Device Model NumberPED-475-25
Device Catalogue NumberPED-475-25
Device Lot NumberA766882
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Initial Date Received by Manufacturer 10/10/2020
Supplement Date Received by Manufacturer10/21/2020
12/18/2020
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date10/30/2020
01/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Patient Weight52
-
-