• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG; PNEUMO SURE HIGH FLOW INSUFFLATOR; INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Model Number 0620040600
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that there was inlet gas pressure issue during the procedure.The procedure was completed successfully.
 
Event or Problem Description
It was reported that there was inlet gas pressure issue during the procedure.The procedure was completed successfully.
 
Additional Manufacturer Narrative
The device was received at the local service facility for investigation.The technical service report is attached (see comm log), and indicates: "calibrate to current specs, upgrade software, 90 day repair warranty".Probable root cause: software malfunction; power supply malfunction; insufflator design; unwanted movement of internal components / wiring; use error; electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge; mio board malfunction; bam board malfunction; lcd assembly failure; manufacturing/ service error.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE HIGH FLOW INSUFFLATOR
Common Device Name
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10673365
Report Number0002936485-2020-00418
Device Sequence Number3312874
Product Code HIF
UDI-Device Identifier07613327063219
UDI-Public07613327063219
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K063367
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0620040600
Device Catalogue Number0620040600
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/17/2020
Supplement Date Received by Manufacturer09/17/2020
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date12/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-