| Model Number |
0620040600 |
| Medical Device Problem Code |
Calibration Problem (2890)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was inlet gas pressure issue during the procedure.The procedure was completed successfully.
|
| |
|
Event or Problem Description
|
|
It was reported that there was inlet gas pressure issue during the procedure.The procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received at the local service facility for investigation.The technical service report is attached (see comm log), and indicates: "calibrate to current specs, upgrade software, 90 day repair warranty".Probable root cause: software malfunction; power supply malfunction; insufflator design; unwanted movement of internal components / wiring; use error; electromagnetic interference (emi) from rf communication, hf surgical instruments, esd, or power surge; mio board malfunction; bam board malfunction; lcd assembly failure; manufacturing/ service error.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
|
| |
|
Search Alerts/Recalls
|