• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012631-59
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual and dimensional analysis was performed on the return device.The reported stent damage was confirmed.The failure to advance and difficulty removing could not be tested as it was based on the circumstances of the procedure.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaints from this lot based on the information provided, the reported difficulties appear to be due to case circumstances.The failure to advance was likely due to interaction with the anatomy.Additionally, during withdrawal, the proximal stent struts likely interacted with the distal end of the introducer sheath causing the struts to flare resulting in resistance with the sheath during removal.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the left common iliac artery with no calcification or tortuosity noted.An 8x59mm omnilink elite balloon expandable stent (bes) was advanced to the aorta without issue, however, there was resistance with the anatomy advancing from the brachial to the common iliac, and the device did not reach the target lesion.The bes was removed with resistance noted, and it was noted that the stent struts were flared.There were no adverse patient effects and no clinically significant delay in the procedure.Another omnilink elite successfully completed the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10673841
Report Number2024168-2020-08561
Device Sequence Number11447524
Product Code NIO
UDI-Device Identifier08717648178528
UDI-Public08717648178528
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model Number1012631-59
Device Catalogue Number1012631-59
Device Lot Number0020741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/18/2020
Initial Report FDA Received Date10/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
-
-