RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Device Displays Incorrect Message (2591); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 13oct2020.
|
|
Event Description
|
The customer reported a unit shut down error and logged a vent-inoperable: machine and proximal pressure sensors failed.There was no patient involvement.
|
|
Manufacturer Narrative
|
G4:22apr2021; b4:26apr2021.The issue was confirmed.The mc to da cable was replaced by the customer and resolved all issues that the v60 was having.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
It is unknown if the issue occurred while in clinical use.No patient or user harm was reported.Multiple good faith efforts were performed to obtain further details regarding the event; however, no response was provided.
|
|
Manufacturer Narrative
|
The device was not being used for medical treatment; no further details were provided regarding the event.No patient or user harm was reported.
|
|