• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LDR MEDICAL ROI-C LORDOTIC IMPLANT 14X17 H8; ROI-C TITANIUM-COATED IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LDR MEDICAL ROI-C LORDOTIC IMPLANT 14X17 H8; ROI-C TITANIUM-COATED IMPLANT SYSTEM Back to Search Results
Catalog Number MC1434P
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3004788213-2020-00166 to 3004788213-2020-00171.
 
Event or Problem Description
It was reported that during the procedure at the c7/t1 level the anchoring plate was difficult to insert the second plate.The cage was found to be fractured, the plate and cage were removed and replaced with alternates of the same size.At the c6/c7 level the first anchoring plate was hindered by the superior endplate of c7 and when the surgeon impacted to insert the plate and cage broke.At the c6 level the anchoring plate was difficult to insert and both the cage and anchoring plate broke.They were removed and replaced with alternates.The surgeon opted to not use anchoring plates at c6 and c7 levels.There were no reported patient impacts.This is report five of six for this event.
 
Event or Problem Description
It was reported that during the procedure at the c7/t1 level the anchoring plate was difficult to insert the second plate.The cage was found to be fractured, the plate and cage were removed and replaced with alternates of the same size.At the c6/c7 level the first anchoring plate was hindered by the superior endplate of c7 and when the surgeon impacted to insert the plate and cage broke.At the c6 level the anchoring plate was difficult to insert and both the cage and anchoring plate broke.They were removed and replaced with alternates.The surgeon opted to not use anchoring plates at c6 and c7 levels.There were no reported patient impacts.This is report five of six for this event.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Summary: the complaint is confirmed via photograph for three roi-c cages and two plates for the failure of fracture during plate insertion.Medical records were not provided for review.Device evaluation: product was not returned, but a photo was provided and used for analysis.In the photo, 6 plates are shown.2 plates exhibit fracture, and one plate exhibits deformation.All 3 cages are shown and exhibit fracture.Potential cause: root cause was unable to be determined.Per the surgeon, the patient had sclerotic bone and poor degeneration which may have contributed to the failures.Dhr review and related actions: per dhr review, the part was likely conforming when it left zimmer biomet control.Device use: this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROI-C LORDOTIC IMPLANT 14X17 H8
Common Device Name
ROI-C TITANIUM-COATED IMPLANT SYSTEM
Manufacturer (Section D)
LDR MEDICAL
quartier europe de l¿ouest
5, rue de berlin
sainte-savine, na 10300
FR  10300
MDR Report Key10674845
Report Number3004788213-2020-00171
Device Sequence Number2948907
Product Code OVE
UDI-Device Identifier03662663016490
UDI-Public(01)03662663016490(17)230401(10)58317
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K151934
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Catalogue NumberMC1434P
Device Lot Number58317
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/15/2020
Supplement Date Received by Manufacturer03/17/2021
Initial Report FDA Received Date10/13/2020
Supplement Report FDA Received Date03/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age66 YR
-
-