Catalog Number PFSR200027
Device Problem Display or Visual Feedback Problem (1184)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2020
Event Type  malfunction  
Event Description
It was reported that during a navio procedure, the camera start to show orange led, and warning message appeared on the screen, telling that camera has no full functionality and reduced field of view (infrared).The surgery was completed by using the same device with a minimum delay (less than 30 minutes).This is the second of four incidents.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the reported device (pn 200027 sn (b)(6) ) used for treatment was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional evaluation could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found similar reports, this issue will continue to be monitored.A relationship, if any, between the subject device and the reported event could not be determined.The reported error message indicates that the illuminator leds have gone bad.The illuminator leds on the camera can go bad over time with use and cause floating "dead zones" in the camera view.This problem is physical in nature in the camera only and it needs to be serviced.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Additional information: d4 g1 mdr reporting contact name and address and phone number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NDI CAMERA POLARIS SPECTRA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10675756
MDR Text Key211500945
Report Number3010266064-2020-01862
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556649541
UDI-Public885556649541
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPFSR200027
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received07/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-