| Model Number |
8253200RF |
| Medical Device Problem Codes |
Sensing Intermittently (1558); Failure to Sense (1559); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis found that the device came in with loose clamps and bent connector pins/cable connector.It was replaced with new clamps and cable.The device was repaired, tested and passed all manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported that the patient interface had an intermittent stimulation when stimulating during the procedure.They would stimulate and there would be no stim then they would stim again and it would work (nothing changed in between).There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d11: section d information references the main component of the system.Other relevant device(s) are: product id: 8253001, serial/lot #: (b)(6).Ubd: , udi#:(b)(6).H6: fdm b17, fdr c20 and fdc d14 applies to the mainframe only.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating that the stimulus probe did not immediately respond.The stimulus was triggered again and it activated.There was momentary delay (less than 1 hour).
|
| |
|