Model Number CLH-SC |
Device Problem
Defective Device (2588)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at olympus service operation repair center, it was found that the examination lamp of the device did not lighted up.Other detailed information was not provided.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) has become aware of another failure that the cooling fan was broken.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.Olympus local service engineer report that the thermal fuse of the device was broken.Omsc surmised that the reported phenomenon was occurred due to the thermal fuse of the device.
|
|