| Model Number |
2128-61-012 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: examination of the returned devices confirmed the reported event.No evidence was found indicating the glo adv straight drill driver malfunctioned.The investigation did not indicate that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a total shoulder replacement, the anterior peg peripheral bit ((b)(4)) broke off midway down the threads of the drill bit, twice/ a different bits.The break left half of the threads in the global advantage at drill driver.We changed apg instruments.A surgical delay of five minutes.
|
| |
|