| Model Number |
190-30-04 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.Concomitant medical products: (cn: 170-32-07, sn: (b)(4)) 36+7 biolox 12/14 head; (cn: 130-32-51, sn: (b)(4)) g1 novation neutral liner.
|
| |
|
Event or Problem Description
|
|
As reported, approximately 34 months postop the initial left tha, this (b)(6) y/o female patient¿s left hip stem was revised due to loosening.The head was removed and the poly for the cup was removed to check if the cup was well fixed.A new novation liner was implanted and the patient receives an exactech monobloc stem.Patient was last known to be in stable condition following the event.Devices to be returned.Rg pending.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (h3) the evaluation of the of the revision reported was likely the result of an insufficient bond between the femoral stem and the bone, which led to aseptic (non-infected) femoral stem loosening.(d11) concomitant devices: (cn: 170-32-07, sn: (b)(6) 36+7 biolox 12/14 head.(cn: 130-32-51, sn: (b)(6) g1 novation neutral liner.
|
| |
|