• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG INTROCAN SAFETY; CATHETER,INTRAVASCULAR,THE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG INTROCAN SAFETY; CATHETER,INTRAVASCULAR,THE Back to Search Results
Catalog Number UNKNOWN
Device Problem Break (1069)
Patient Problem Needle Stick/Puncture (2462)
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.Additional attempts to receive more information on the reported event are being made.A follow-up report will be submitted when the results of the investigation are available.
 
Event Description
As reported by user facility: as rn tried to advance, the plastic part tore.Entire piece was retrieved and only harm was that patient had to be stuck again.
 
Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The actual device involved in the reported incident was not returned, however a photo was provided for evaluation.The photo shows that the needle has pierced the catheter after the device was used.The most likely cause for this is if the needle was re-inserted.The ifu for introcan safety products does state, "after withdrawal, do not reintroduce the steel needle into the catheter, as the latter may be cut off,".If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTROCAN SAFETY
Type of Device
CATHETER,INTRAVASCULAR,THE
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, hessen 34212
GM  34212
MDR Report Key10677810
MDR Text Key211304631
Report Number9610825-2020-00252
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
PMA/PMN Number
K020785
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/29/2020
Initial Date FDA Received10/14/2020
Supplement Dates Manufacturer Received09/29/2020
Supplement Dates FDA Received11/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-