| Model Number |
MSERIES BI-PHASIC |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device failed self test.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.The customer's report was verified and attributed to a faulty transient voltage suppressor diode on the bridge board.The bridge board was replaced to remedy the report.The device was re-certified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|