• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 4580 FMS DUO+ PUMP/SHAVER COMBO -NS; DISTENSION UNIT, FLUID, ARTHROSCOPIC Back to Search Results
Model Number 284580
Device Problem Break (1069)
Patient Problem Insufficient Information (4580)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Udi: (b)(4).Investigation summary: the device was received at the service center and evaluated.There was no allegation of malfunction against the device from the customer, however, defects were found with the device during service evaluation.Per service manual operational and diagnostic, the irrigation cover and two hinges of the device were found to be broken; therefore, this complaint can be confirmed.The left lexan safety cover and hinges were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.With the available information, we cannot determine the root cause of the identified problems.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 10/02/2018 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Initial reporter phone number: (b)(6).
 
Event Description
It was reported by the affiliate via service request that the 4580 fms duo+ pump/shaver combo -ns had irrigation cover and 2 damaged hinges.Additional information could not be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4580 FMS DUO+ PUMP/SHAVER COMBO -NS
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key10680030
MDR Text Key213697521
Report Number1221934-2020-02998
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020607
UDI-Public10886705020607
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number284580
Device Catalogue Number284580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2020
Date Manufacturer Received10/07/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-