Model Number 6260 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.This event is related to mdr # 1810909-2020-00546.
|
|
Event Description
|
The customer reported that her contour next link 2.4 meter was transferring the blood glucose readings twice to her connected insulin pump.The customer provided the example where blood glucose readings of 168 mg/dl and 268 mg/dl were obtained and the insulin pump delivered 1.6 units of insulin and 4.6 units of insulin respectively.The customer stated that the insulin was received based on the first reading.No insulin was received based on the duplicate reading sent to the insulin pump.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter kit and test strips were sent to the customer.
|
|
Manufacturer Narrative
|
The customer returned the suspected contour next link 2.4 meter for evaluation.No test strips were returned.The meter service mode was run and no anomalies were found.The returned meter was tested with the in-house contour next test strips and in-house breeze 2 control solution to simulate as blood.Which gave a satisfactory performance.The returned meter was connected with a reference insulin pump.The simulated blood glucose readings were properly transferred to the insulin pump.The meter did not send duplicated results to the insulin pump.
|
|
Search Alerts/Recalls
|