Model Number 8900-2106-01
Medical Device Problem Code Arcing of Electrodes (2289)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), an arc was heard from the electrode pads.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation evaluated the device and the device performed to specification.Inspection of the pads did find discoloration on the conductive plates indicating a discharge occurred.Review of the device clinical data showed one shock event for the reported event date and the output was within specification for the measured impedance.There are no abnormalities or errors observed in the clinical data.Electrode labeling calls out the importance of good placement on the patient and provide instructions for proper electrode application technique.Poor adherence and /or air under the electrodes can lead to the possibility of arcing and skin burns.Good skin preparation does not guarantee a burn will not occur and burns from defibrillation is an expected risk.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRO-PADZ LIQUID GEL RADIOLUCENT MFE, 1 SET
Common Device Name
ELECTRODE
Manufacturer (Section D)
BIO-DETEK INCORPORATED
525 narragansett park drive
pawtucket RI 02861
MDR Report Key10680245
Report Number1218058-2020-00076
Device Sequence Number11536651
Product Code MKJ
UDI-Device Identifier00847946016401
UDI-Public00847946016401
Combination Product (Y/N)N
Initial Reporter CountryAU
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date10/26/2020
Device Model Number8900-2106-01
Device Catalogue Number8900-2106-01
Device Lot Number4319
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/23/2020
Supplement Date Received by Manufacturer09/23/2020
Initial Report FDA Received Date10/14/2020
Supplement Report FDA Received Date11/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-