It was reported that the port cap of a 3000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was partially detached.The issue was further described as, ¿the administration port cap had started to come off on a home pn bag." the issue was identified prior to patient use.There was no report of patient injury, or medical intervention associated with this event.No additional information is available.
|
Additional information was added to h3, h4 and h6.H4: the lot was manufactured from january 09, 2020 - january 13, 2020.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The photograph was evaluated which observed that the spike port cap was almost completely cracked off from the spike port cap.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|