| Model Number |
9-ASD-012 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
09/11/2020
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020, a 12mm amplatzer septal occluder was selected for implant.During deployment, a cobra-deformation was noted.The device was pull back into the sheath and detached into the svcand the deformation was still noted.The occluder was snared though a large sheath.A 18mm amplatzer p.I.Muscular vsd occluder was implanted.The patient was transferred to the intensive care unit (icu) and was unstable.A few hours post procedure the patient became hemodynamically unstable.It was reported that the muscular vsd occluder had embolized into the right ventricle and the patient expired on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of deformation during deployment could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note, per the instructions for use artmt100116885 rev.A, "the amplatzer¿ septal occluder is a percutaneous, transcatheter atrial septal defect closure device intended for the occlusion of atrial septal defects (asds) in secundum position or patients who have undergone a fenestrated fontan procedure and who now require closure of the fenestration.".
|
| |
|
Search Alerts/Recalls
|