• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO BONE PIN 4.0MMX152MM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO BONE PIN 4.0MMX152MM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB00031
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/16/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a navio tka procedure, when drilling the box, the surgeon hit the femoral bone pin.This damaged the bone pin (bent) making it unusable (it was no longer required either).It did not negatively affect the case (there were no delays and no injuries).No other complications were reported.
 
Additional Manufacturer Narrative
The navio bone pin, part rob00031, used for treatment was not made available to the designated complaint unit for evaluation thus, a visual and functional evaluation could not be performed.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, the most likely cause of this event is improper handling of the part.Based on the investigation, no further containment or corrective action is recommended or required at this time.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO BONE PIN 4.0MMX152MM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key10680633
Report Number3010266064-2020-01867
Device Sequence Number303786
Product Code OLO
UDI-Device Identifier00885556685310
UDI-Public00885556685310
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K191223
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberROB00031
Device Catalogue NumberROB00031
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 09/16/2020
Supplement Date Received by Manufacturer06/09/2021
Initial Report FDA Received Date10/14/2020
Supplement Report FDA Received Date06/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-