Intera is aware that the tapered catheter is being used off-label with the medtronic pump for hepatic artery infusion.The surgical staff disposed of the catheter, so it was unavailable for physical evaluation.The lot used in the surgery was unknown.According to intera sales records, only two lots of catheters were sold to the reporter's institution, lot j32r96 and j58x05.The lot history records for both lots were reviewed and no nonconformances or deviations were noted in the records.No other reports of missing beads/rings have been reported to intera.For reference, the beads/rings are located at the end of the catheter.The surgeon may cut the end of the catheter when prepping for surgery and it may be possible to cut the beads/rings.Since no physical evaluation was able to be performed, it is undetermined if the catheter was cut in a manner that would have cut the bead.Since no other complaints have been reported and no product defect trend is detected, this complaint will be considered closed with no further corrective action.If additional complaint(s) are reported, corrective action will be considered.
|