• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII PS INSERT SZ 3-4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII PS INSERT SZ 3-4 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71420818
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/16/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the gii ps insert sz 3-4 11mm is not locking properly.It is unknown when was this malfunction discovered.There was no delay and a s&n back up device was available.No injuries reported at this time.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned insert shows minor damage to the lock detail, more than likely from trying to force the insert to lock into the baseplate.A dimensional evaluation could not be completed due to the damage to the lock detail.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Factors and/or some potential probable causes that could contribute to the reported event have been identified as overuse or excessive pressure on the joint, injury, infection and/or patient condition.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII PS INSERT SZ 3-4 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10681166
Report Number1020279-2020-05377
Device Sequence Number2991479
Product Code JWH
UDI-Device Identifier03596010209580
UDI-Public03596010209580
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K951987
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71420818
Device Catalogue Number71420818
Device Lot NumberJ1902394
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2021
Initial Date Received by Manufacturer 09/16/2020
Supplement Date Received by Manufacturer06/24/2021
Initial Report FDA Received Date10/14/2020
Supplement Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight81
-
-