| Model Number |
357494 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A passeo-18 balloon catheter was chosen for the treatment of a mildly calcified lesion.The balloon ruptured at 5 atm.The balloon could be withdrawn and another balloon was used.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed that the balloon has burst longitudinal over a length of about 36 mm.Microscopic inspection of the balloon surface showed several scratches in close vicinity of the tear.It seems likely that the tear in the balloon was caused by a hard, sharpedged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause was determined.The root cause is most likely related to the patients anatomy.
|
| |
|
Search Alerts/Recalls
|