Model Number V60 |
Device Problem
Failure to Read Input Signal (1581)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 14oct2020.
|
|
Event Description
|
The customer reported that the touch screen is not sensitive and unresponsive.The customer contacted product support and requested for service repair.The customer reported that the unit was in use on patient, but there was no patient harm reported.
|
|
Manufacturer Narrative
|
G4: 14oct2020; b4: 27oct2020.Per service engineer the issue was discovered during testing.There was no patient involvement.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4: 12oct2020.B4: 23mar2021.The authorized service personnel (asp) replaced the rp-touch screen and rp-overlay, power switch,eng,v60,start ventilator, everything returned to normal.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|