• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TF ULTRA 2 UBRAID AND NDLS 5.5MM TI; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TF ULTRA 2 UBRAID AND NDLS 5.5MM TI; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202618
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Information (3190)
Date of Event 09/25/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an acl procedure, the anchor of the twinfix failed at the time of placing in the patella.The proper practice of placing the anchor was followed but it remained outside the patella and additional maneuvers had to be performed to finish inserting it since the insertion handle was rolled and did not continue to insert the anchor, so it was outside the patella.Procedure was successfully completed with the same device using additional tools.No significant delay and no patient complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A clinical evaluation was conducted and found without the requested clinical information a thorough medical investigation cannot be rendered.It is unknown what ¿additional maneuvers¿ were performed to complete the procedure.Per complaint details, the additional maneuvers were performed to finish inserting the twinfix.The patient impact could not be determined based on the limited information provided.Should any additional clinical information be provided this complaint will be re-evaluated.No further clinical/medical assessment is warranted at this time.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TF ULTRA 2 UBRAID AND NDLS 5.5MM TI
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10681278
Report Number1219602-2020-01583
Device Sequence Number2962268
Product Code MBI
UDI-Device Identifier03596010656131
UDI-Public03596010656131
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K100159
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/12/2024
Device Model Number72202618
Device Catalogue Number72202618
Device Lot Number2044858
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/28/2020
Supplement Date Received by Manufacturer04/19/2021
Initial Report FDA Received Date10/14/2020
Supplement Report FDA Received Date04/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-