• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. CNNLTD 5.5 PLYSCRW 6.5X45; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. CNNLTD 5.5 PLYSCRW 6.5X45; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3505-6545
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event or Problem Description
It was reported that during the procedure the surgeon wanted to revise the screw to change the trajectory.The screw was removed using the removal tool, but then broke after it was removed from the body.An alternate screw of the same size was used to complete the case.There were no reported patient impacts.
 
Additional Manufacturer Narrative
Corrected information in d10 and h3.Additional information in h6: method, results, and conclusions.This follow-up report is being submitted to relay additional information).The complaint is unconfirmed for one (1) of one (1) path finder nxt implant (part number 3505-6545) for the failure of fracture, but confirmed for the failure of disassembly.Device evaluation: the item returned matches the item in the complaint.Upon visual inspection the tulip head was detached from the screw.The complaint is confirmed.Complaint history search: a complaint history search was performed.Dhr review and related actions per dhr review, the part was likely conforming when it left zimmer biomet control.No actions required.This event is not related to any current actions or recalls or product holds.Device use this device is used for treatment.Potential cause with the given information, the definitive cause cannot be determined.Typically, the cause is attributed to locking the closure top into place with the pedicle screw and rod, then subsequent loosening of the closure top for adjustment.The splines deform during locking in order to lock the tulip into position and prevent it from moving about the screw shaft in situ.The deformation on the splines also prevents the screw from being reused, as it allows the tulip to disassemble from the shaft once the closure top is loosened or removed.
 
Event or Problem Description
It was reported that during the procedure the surgeon wanted to revise the screw to change the trajectory.The screw was removed using the removal tool, but then broke after it was removed from the body.An alternate screw of the same size was used to complete the case.There were no reported patient impacts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNNLTD 5.5 PLYSCRW 6.5X45
Common Device Name
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
MDR Report Key10681545
Report Number3012447612-2020-00573
Device Sequence Number16286532
Product Code NKB
UDI-Device Identifier00889024335707
UDI-Public(01)00889024335707(10)AAL
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K132884
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3505-6545
Device Lot NumberAAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/16/2020
Supplement Date Received by Manufacturer12/11/2020
Initial Report FDA Received Date10/14/2020
Supplement Report FDA Received Date12/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No