Model Number 204-24-11
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/01/2009
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that a patient experienced a right knee revision surgery on (b)(6) 2009, due to unreported reasons.Initial surgery date was not reported.No additional information about the event or patient has been provided.These devices are used for treatment not diagnosis.It is found in a review of the labeling and ifu 700-096-004 rev.N, it is known that device specific risks consist of: fracture, migration, loosening, subluxation, or dislocation of the prothesis or any of its components, any of which may require a second surgical intervention or revision.As part of the preoperative assessment, the surgeon must ensure that no biological, biomechanical, or other factors exist that might adversely affect the surgery and/or postoperative period.Revisions or surgical interventions are a known complication found in joint replacements.All patients should be instructed on the limitations of the prosthesis and the possibility of subsequent surgery.The patient should be cautioned to monitor activities and protect the replaced joint from unreasonable stresses and follow the written instructions of the physician with respect to follow-up care and treatment.The surgeon must be fully knowledgeable about all aspects of the surgical technique to use the implants in accordance with the indications and contraindications and be trained according to the proper use of the system instrumentation and implants.It is a known complication that a patient's age, weight, or activity level would cause the surgeon to expect early failure of the system.As noted in "why are total knee arthroplasties failing today - has anything changed after 10 years?", the top five reasons for total knee arthroplasty revision/surgical intervention account for approximately 85% of the reasons that drive revision/surgical intervention.The failure mechanisms are as follows: 39.9% for loosening; 27.4% for infection; 7.35% for instability; 4.7% for periprosthetic fracture; and 4.5 % for arthrofibrosis.In early revisions (less than two years) infection was found to be the most common failure mechanism and late revisions show aseptic loosening as the most common reason.1 1.Sharkey, p.L.(2014, september).Why are total knee arthroplasties failing today - has anything changed after 10 years?; the journal of arthroplasty, 29.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The most likely cause cannot be determined, due to the lack of information.There has been no patient information provided; therefore, the patient risk/clinical factors cannot be assessed.This event is being reported as part of a large group of mdr reportable events (386) received from the hospital for specialty surgery (hss).Exactech engineering became aware of unreported revisions of exactech knee product via a published abstract and poster (paper 640 what happened? extreme delamination and oxidation in modern day compression molded polyethylene brian t.Nickel, md, new york, ny lydia weitzler, ms, new york, ny douglas e.Padgett, md, new york, ny timothy m.Wright, phd, new york, ny.Exactech engineering requested access to the hss's details regarding the data used for the publication.The knee engineering team received raw data from hss documenting 386 mdr reportable events (revisions) that occurred at hss from 2001 through january 2018.Hss did not report these events individually to exactech at the time of the revisions.Exactech engineering forwarded the information to the exactech complaint handling unit.Each individual event has been added to the complaint data and investigated.Capa (b)(4) was opened to address this issue and to investigate and report these events.As part of the capa, a risk assessment was conducted to compare the rate of knee revisions before and after adding the additional revisions from hss to the complaint data.The risk assessment concluded that the overall occurrence rate remains at a rating of very low (1) per exactech's risk management procedure.The overall risk profile of the device has not changed.
 
Event or Problem Description
It was reported that a patient experienced a right knee revision surgery on (b)(6) 2009, due to unreported reasons.Initial surgery date was not reported.No additional information about the event or patient has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 4, 11MM
Common Device Name
PS TIBIAL INSERTS SZ 4, 11MM
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
MDR Report Key10681639
Report Number1038671-2020-10070
Device Sequence Number11540752
Product Code JWH
UDI-Device Identifier10885862048394
UDI-Public10885862048394
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2009
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,literatur
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/05/2011
Device Model Number204-24-11
Device Catalogue Number204-24-11
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2018
Initial Report FDA Received Date10/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Disability;
-
-