Catalog Number UNKNOWN
Device Problems Inadequate Filtration Process (2308); Obstruction of Flow (2423); Structural Problem (2506); Insufficient Information (3190)
Patient Problems Death (1802); Internal Organ Perforation (1987); Thrombus (2101); Perforation of Vessels (2135)
Event Type  Death  
Manufacturer Narrative
(b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.The reported allegations have been investigated based on the information provided to date.Unknown if the reported death is directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Cook medical will continue to monitor for similar events.
 
Event Description
It is alleged that "[pt] received a cook celect filter on (b)(6) 2013." patient is deceased.Hospital and medical records have been requested but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2013 via the right internal jugular vein due to deep vein thrombosis (dvt) and pulmonary embolism (pe).Allegations include: organ perforation and death.It is further alleged "that the filter contributed to the [decedent's] heart failure and death.He contends that the decedent could not walk more than 30 minutes because of weakness in [patient's] legs and shortness of breath," as well as post-implant pe.A certificate of death dated on (b)(6) 2018 has been received without a listed cause of death.On (b)(6) 2017, report from ct (computed tomography): "tumor thrombus / clot burden seen throughout the course of the ivc and central mesenteric veins.An ivc filter tine projects outside the ivc lumen medially into the aorta." "worsening groundglass opacities in the visualized rml and right lower lobe likely sequela of pulmonary infarction from known extensive pulmonary emboli.".
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: pulmonary embolism (pe), vena cava (vc) / organ / aorta perforation, inferior vena cava (ivc) thrombus, heart failure, dyspnea, physical limitations.New pe as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: pulmonary embolism.Filter interacts with ivc wall, e.G.Penetration / perforation / embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Ivc occlusion / thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Unknown if the reported heart failure, dyspnea, and physical limitations are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key10682160
MDR Text Key211421853
Report Number3002808486-2020-00938
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/06/2020
Initial Date FDA Received10/15/2020
Supplement Dates Manufacturer Received11/10/2020
Supplement Dates FDA Received12/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Death;
-
-