MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETLW1610C156E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Arrest (1762); Death (1802); Low Blood Pressure/ Hypotension (1914)
Date of Event 09/23/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant iis stent graft system was implanted in a patient for the endovascular treatment of a left common iliac aneurysm and ectatic right common iliac.It was reported during the index procedure after the stent graft was deployed, final angio showed a good result and the patient was taken to sicu post-op.It was reported two hours post transfer the patient became hypotensive and coded.The patient was transferred to the cath lab where cpr was in process.An angio showed a ruptured right external iliac, it was unclear were the rupture was so an etlw1610c156e (b)(4) was implanted to cover the right hypogastric and extended into the external iliac.The bleeding was gotten under control however the patient had lost too much blood and had been getting cpr and resuscitated for an hour at this time.The physician decided to stop and pronounced the patient deceased.Per the physician the cause of the death was due to the ruptured external iliac that was not visualized on the completion angiogram.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
IE 
MDR Report Key10682748
Report Number9612164-2020-03932
Device Sequence Number14243614
Product Code MIH
UDI-Device Identifier00643169467637
UDI-Public00643169467637
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/22/2021
Device Model NumberETLW1610C156E
Device Catalogue NumberETLW1610C156E
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2020
Initial Report FDA Received Date10/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Required Intervention;
Patient Age78 YR
Patient Weight84
-
-