Model Number 110-150
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 02/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture baker grade iii is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation: capsular contracture baker grade iii.
 
Event or Problem Description
Healthcare professional reported for right side, capsular contracture baker grade iii.Device has been explanted.
 
Event or Problem Description
Healthcare professional reported for right side, capsular contracture baker grade iii.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: analysis of the returned device identified: weight to the spec, bubbles in the gel, wear abrasion and crease fold.Cloudy and voids in the gel observed after autoclave cycle.Base on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 110 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (AUSTIN)
12331-a riata trace parkway
building 3
austin TX 78727
MDR Report Key10682816
Report Number9617229-2020-10648
Device Sequence Number436743
Product Code FTR
UDI-Device Identifier10888628002173
UDI-Public10888628002173
Combination Product (Y/N)N
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number110-150
Device Catalogue Number110-150
Device Lot Number3124792
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Initial Date Received by Manufacturer 02/05/2020
Supplement Date Received by Manufacturer01/14/2021
Initial Report FDA Received Date10/15/2020
Supplement Report FDA Received Date02/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
-
-