Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the provisional broke during trialing.No further information is available at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay corrected and additional information.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows that the trial provisional was fractured and all pieces were returned.The device also exhibits signs of usage.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause is considered to be determined as normal wear and tear from repeated use during the instrument's field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|