• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH- AT ARTIS Q.ZEN CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH- AT ARTIS Q.ZEN CEILING; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10848354
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Unspecified Tissue Injury (4559)
Date of Event 10/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event or Problem Description
It was reported to siemens that an adverse event occurred while operating the artis q.Zen ceiling system.The user reported that the table head end holder slid from the table and landed on the operators toes.Information was provided that the operators foot is sore, however, it is not clear if any medical treatment was needed.There is no report of impact to the state of health of any patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Additional Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause could not be determined.The investigation could not provide a clear result as there are two possibilities for the cause of the error.The first possibility is that the problem was caused by a loose screw.The second possibility is that the pluggable handle was not correctly fastened by the user beforehand.Both cases are plausible, but it is no longer possible to determine them beyond doubt as the technical conditions on site have changed in the meantime.The affected part has been checked and secured by the local service organization.The error has not been reported again.The system is in proper condition and functions as specified.A possible general error could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS Q.ZEN CEILING
Common Device Name
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH- AT
siemensstrasse 1
forchheim, germany 91301
GM  91301
MDR Report Key10683130
Report Number3004977335-2020-50012
Device Sequence Number3450075
Product Code OWB
UDI-Device Identifier04056869010021
UDI-Public04056869010021
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K181407
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Remedial Action Repair,Inspection
Type of Report Initial,Followup
Report Date (Section B) 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number10848354
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/07/2020
Supplement Date Received by Manufacturer12/07/2020
Initial Report FDA Received Date10/15/2020
Supplement Report FDA Received Date12/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-