| Model Number |
SEE H10 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
09/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Reference exemption number e2017029.An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The failure root cause cannot be determined due to the device not being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported a twist drill broke during use.A portion remains implanted.The reported rpms exceeded the maximum value found in the ifu.Per ifu 91-600-02-00 for the twist drill "operate the twist drills with no more than the maximum recommended handpiece speed setting of 2,500 rpm.".
|
| |
|
Additional Manufacturer Narrative
|
|
E4: multiple mdr reports were filed for this event, please see associated report: mdr 3933020000-2020-8014.
|
| |
|
Search Alerts/Recalls
|