• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number FX-410360
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.The event of "infection (unknown onset)" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: infection.Further investigation results: with the information collected during the investigation, there is enough evidence to support that device serial number (b)(4) was manufactured under controlled conditions in accordance with allergan procedures.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run (b)(4) is not related to any additional complaint infection record reported as of today.During the trend review of all infection complaints for gel breast implants for the period of apr-18 through mar-20, was noted an outlier in sep -18.An additional analysis was performed to determine any pattern or any similarities relation between pr¿s involved.It was found that some primary packaging operators were involved in more than one occasion, however the people were trained in the corresponding procedure.Also, calibrations, maintenance and validations of the equipment¿s involved in more than one occasion were reviewed and it was determined that they were performed on time and met specifications.In addition, all the records involved in this month were reviewed and no issues were found associated to either event.Given the fact that the cause of the infection cannot be specifically associated to manufacturing process, and there is not an adverse trend for this type of event no corrective action is deemed necessary at this time.
 
Event or Problem Description
Patient reported "implant has been removed due to infection." this record is for the right side.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key10683633
Report Number9617229-2020-07260
Device Sequence Number14162228
Product Code FTR
UDI-Device Identifier10888628002975
UDI-Public10888628002975
Combination Product (Y/N)N
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/09/2020
Device Model NumberFX-410360
Device Catalogue NumberFX-410360
Device Lot Number2702452
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2020
Initial Report FDA Received Date10/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-