| Model Number |
TRANSDUCER X8-2T |
| Medical Device Problem Codes |
Positioning Failure (1158); Mechanical Jam (2983)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Return of the suspect transducer is anticipated.Evaluation of the transducer will be included in a follow up report upon its return and investigation completion.
|
| |
|
Event or Problem Description
|
|
A customer reported an x8-2t model transducer had an articulation issue preventing tip motion.There was no injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the transducer confirmed the articulation issue as described by the customer.Functional testing of the device noted failure of the image, black box, and articulation tests.Visual inspection noted a hole in the tip shell and inferior beading.Destructive testing identified an insufficient solder joint allowed the steering cable to be separated from the rack actuator which resulted in the loss of articulation.The failed transducer was replaced by the philips service engineer at the customer site and no similar issues have been reported post repair.
|
| |
|
Event or Problem Description
|
|
A customer reported an x8-2t model transducer had an articulation issue preventing tip motion.There was no injury associated with this event.
|
| |
|