• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550225-15
Medical Device Problem Codes Entrapment of Device (1212); Improper or Incorrect Procedure or Method (2017); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.It should be noted that the xience sierra everolimus eluting coronary stent system instruction for use states: applying excessive force to the delivery system can potentially result in loss or damage to the stent and/or delivery system components.The investigation determined the reported difficulties and treatment appear to be related to circumstances of the procedure as it is likely the device interacted with the previously implanted stent during advancement causing the reported failure to advance.A projecting strut from the previously implanted stent entrapped the device during removal causing the reported stent entrapment.Force was reportedly applied due to the entrapment causing the reported stent dislodgement and subsequent treatment.There is no indication of a product quality issue with respect to manufacture, design or labeling.Na.
 
Event or Problem Description
It was reported that the procedure was performed to treat a distal left anterior descending artery.During advancement of the 2.25x15mm xience sierra stent delivery system (sds), the stent became caught with a projecting strut of a previously implanted unspecified stent.The sds was pushed but it failed to cross.Strong resistance was felt during removal.Force was applied and the stent dislodged.The dislodged xience sierra stent was retrieved with a snare device.No other stent was implanted.The procedure was completed using a balloon dilatation catheter.There was no adverse patient sequela reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10684391
Report Number2024168-2020-08624
Device Sequence Number16286521
Product Code NIQ
UDI-Device Identifier08717648227011
UDI-Public08717648227011
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2021
Device Model Number1550225-15
Device Catalogue Number1550225-15
Device Lot Number8062741
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2020
Initial Report FDA Received Date10/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-