| Model Number |
PICC-142 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that the hydropicc replaced an unknown picc device (not a hydropicc), which migrated into the patient's axillary vein, occluding that vein and causing a clot and phlebitis in his right arm; the picc had been implanted for 4 days.The hydropicc was replaced on the same day with another hydropicc into the left brachial vein, and secured with a 5fr securacath.No issues were noted with the replacement procedure.A review of lot #09032001, including the associated lot release data, was performed and did not reveal any notations, comments, or issues, which could be attributed to the cause of the reported problem.The complainant has indicated that the suspect device is being returned for analysis; however, the device has not been received yet.
|
| |
|
Event or Problem Description
|
|
It was reported that, 1 day after being implanted, the hydropicc was unable to be flushed, and that a guidewire could not be advanced through the entire length of the catheter.The catheter was explanted where a knot was observed at the distal end of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Although several attempts were made to retrieve the suspect device, it has not been returned for evaluation.Therefore, the reported complaint could not be confirmed and a root cause could not be determined.
|
| |
|