| Catalog Number |
07.02072.001 |
| Medical Device Problem Codes |
Fracture (1260); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The complaint is confirmed for 1 returned vitality curved lumbar lenke for the reported failure of bending/fracturing.Visual inspection revealed that the tip of the device has broken off.Per dhr review, the part was likely conforming when it left zimmer biomet control.No actions required.This event is not related to any previous actions.No holds or recalls are associated with this part number.The damage to the device likely occurred during regular use and was a result of normal wear.However, a definitive root cause could not be determined.If any further information is found which would change or alter any conclusions, or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery, the probe became bent.While attempting to straighten the bent probe on the back table in the or, the tip broke off.There was no impact on the patient.
|
| |
|