| Model Number |
5571 |
| Medical Device Problem Code |
Peeled/Delaminated (1454)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturing record was reviewed.There were no nonconformance related to this lot, therefore, supporting the device met material, assembly and performance specifications.A follow-up report will be issued after the investigation is complete.
|
| |
|
Event or Problem Description
|
|
As reported: started to load guideliner v3 onto wire when "hypo tube" caught on valve.Physician noticed that with each push, the back dark blue plastic piece that extends from "tube" onto wire, started to come off wire.Device never made it into body, or guide, so no intervention needed to retrieve anything.No harm due to device issue.
|
| |
|
Additional Manufacturer Narrative
|
|
Guideliner was returned to vsi for evaluation.The guideliner pushrod was separating from the proximal half-pipe.The manufacturing records were reviewed, and there were no nonconformances related to this lot, therefore, supporting the device met material, assembly, and performance specifications.The event description states the guideliner broke apart.The underside of the guideliner had the push rod exposed through the pebax material.No other damage was noted.An internal corrective action was initiated to further investigate.Per the investigation, no probable cause was determined, no root cause could be established.
|
| |
|