Product complaint#: (b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date, a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: what was the initial procedure? unknown.What is the event day and procedure date? unknown.What is the lot number? qcbher.Could you please confirm how many devices present the issue (pull off suture needle intra op) during this procedure? 2.
|
|
(b)(4).Date sent to the fda: 12/1/2020.Additional information: d9, h6.A manufacturing record evaluation was performed for the finished device batch qcbher, 18515g55 and no non-conformances were identified.Additional h3 investigation summary: it was reported that suture pull-off during dispense from the needle.Suturing could not be done with the needle-suture.It was a control release needle.Seven unopened samples and an empty opened box of product code 18515, lot qcbher were received for evaluation.During the visual inspection of seven unopened samples, no defects were found on the packages.The samples were opened, and swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.The device performed without any defect noted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|