MAUDE Adverse Event Report: VYAIRE MEDICAL VMAX ENCORE 22D; CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL VMAX ENCORE 22D; CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION Back to Search Results
Model Number VMAX 22D E
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Vyaire complaint number (b)(4).Any additional information provided by the customer will be included in a follow-up report.
 
Event or Problem Description
A customer contacted vyaire medical to report that they noted the fan inside the dell pc was, "whining and smelled hot." upon replacing the pc, they started getting gas not connected warnings even though they were connected to the wall o2.Customer was advised to put the pc back to the original system, and replace the sol cable from the system in order to resolve the communication problem.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VMAX ENCORE 22D
Common Device Name
CALCULATOR, PREDICTED VALUES, PULMONARY FUNCTION
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa IL 60045
MDR Report Key10686886
Report Number2021710-2020-12725
Device Sequence Number19635312
Product Code BTY
UDI-Device Identifier10846446021417
UDI-Public(01)10846446021417(11)20091016
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K942211
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVMAX 22D E
Device Catalogue Number777405-101
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/23/2020
Initial Report FDA Received Date10/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-