| Model Number |
VMAX 22D E |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Vyaire complaint number (b)(4).Any additional information provided by the customer will be included in a follow-up report.
|
| |
|
Event or Problem Description
|
|
A customer contacted vyaire medical to report that they noted the fan inside the dell pc was, "whining and smelled hot." upon replacing the pc, they started getting gas not connected warnings even though they were connected to the wall o2.Customer was advised to put the pc back to the original system, and replace the sol cable from the system in order to resolve the communication problem.There was no patient involvement.
|
| |
|