Model Number MMT-551NAS |
Device Problems
Mechanical Problem (1384); No Apparent Adverse Event (3189)
|
Patient Problem
Skin Inflammation (2443)
|
Event Date 10/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
Information received by medtronic indicated that insulin pump had loud during rewind since it was struggling to complete.Customer had skin issue so had to stop using years ago so transmitter would no longer hold a charge.No harm requiring medical intervention was reported.The customer was assisted with troubleshooting.The device will be returned for analysis.
|
|
Manufacturer Narrative
|
Device monitored for several days.Device received with normal operating current.Device passed displacement test, rewind and self test.Device passed off no power test.No unexpected rewind anomalies noted.No unexpected low battery alarm anomalies noted.Device received with cracked belt clip slot and cracked reservoir tube lip.
|
|