SMITH & NEPHEW, INC. TANDEM STR THREADED HNDLE; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Model Number |
71362804 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the handle broke off.Patient injuries were not reported.It is unknown when the event occurred, how the procedure finished and if there were a surgical delay.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the tandem str threaded hndle is broken in two pieces at the handle.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|