• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US HEALIX AWL -NS; BONE AWL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US HEALIX AWL -NS; BONE AWL Back to Search Results
Model Number 222223
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Udi: (b)(4).
 
Event or Problem Description
It was reported by the sales rep via phone that the tip of the healix awl is bent.It is unknown when the failure was found.The device was discarded by the staff.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (1205001), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX AWL -NS
Common Device Name
BONE AWL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10687491
Report Number1221934-2020-03022
Device Sequence Number19610495
Product Code LXH
UDI-Device Identifier10886705007707
UDI-Public10886705007707
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number222223
Device Catalogue Number222223
Device Lot Number1205001
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/28/2020
Supplement Date Received by Manufacturer11/01/2020
Initial Report FDA Received Date10/15/2020
Supplement Report FDA Received Date11/02/2020
Patient Sequence Number1
-
-